Agilent 1290 Infinity Binary Pump User Manual 53
Using the Pump
4
Setting up the Pump with the Instrument Control Interface
Timetable Settings
Use the Timetable to program changes in the pump parameters during the
analysis by entering a time in the Time field and appropriate values in the
following fields of the timetable. Changes in flow rate occur linearly from
either time zero or the time of the last defined change; other parameters
change instantaneously at the time defined in the timetable.
Show Advanced Timetable toggles the timetable display between standard mode
and advanced mode.
The following parameters can be changed:
Change Contacts
•Change Flow
Change Max. Pressure Limit
Change Solvent Composition - You can only use solvents, which have been
enabled in the solvents section.
Function centric view - This checkbox allows you displaying parameter
changes instead of a time table.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals