64 Agilent 1290 Infinity Binary Pump User Manual
5 How to Optimize the Performance of Your Module
How to Configure the Optimum Delay Volume
Solvent packages created by the pump may persist until the solvent reaches
the detector flow cell. Absorption fluctuations can then show up as baseline
noise, also referred to as mixing noise. Applications like impurity
quantitation or lowest level compound detection require minimizing this
noise. The 380 µL Jet Weaver strongly improves mixing and therefore
reduces baseline noise and improves sensitivity in detection. Patented
Agilent microfluidic technology offers high mixing performance at a low
internal volume of 380 µL.