Remote Operation
M370430-01 Rev B 5-25
5
Multichannel Address Setting
The power supply multichannel address can be set to any address
between 1 to 30. All units that are connected together via the RS-232 or
RS-485 connector must have a unique multichannel address.
To set the address:
1. Select the appropriate communication port as described
“Selecting the Appropriate Communication Port” on page
5-24.
If more than one unit has the same multichannel address when
connected through the RS-485 connector, then collisions are
possible, resulting in garbled responses to queries.
2. Press the rotary Adjust/Enter control.
is displayed on the output voltage display.
3. Turn the rotary Adjust/Enter control to select the desired
multichannel address between 1 to 30.
4. Press the rotary Adjust/Enter control to commit the new
address.