Local Operation 
 
M370430-01  Rev B  3-31 
Defining the UVP Set Point 
To define the UVP set point: 
1.  Turn the power supply On. 
2.  Set the output to the desired voltage. 
3.  Turn the 9-position mode control to the PRT position. 
 is displayed. 
4.  Turn the rotary Adjust/Enter control until   is 
displayed. 
5.  Press the rotary Adjust/Enter control to select the UVP for 
setting up. 
6.  Set up the lower limit protection using the rotary Adjust/Enter 
control. 
Important:  The UVP range is variable from 0 volts to OVP
SET
 -10%. 
 
7.  Once the desired value has been set, press the rotary 
Adjust/Enter control to commit the setting. 
See “Coarse and Fine Adjustment Modes” on page 3-3. 
The SCPI command (s) for these instructions are: 
[[:]SOURce[<channel>]]:VOLTage:PROTection:UNDer[:LEVel] 
 
Over Current Protection (OCP) 
The OCP alarm is a firmware alarm that is only configurable using the 
SCPI interface. The OCP triggers if the output current exceeds the OCP 
set point. Because the OCP alarm is a firmware based alarm it has a 
latency before detection of the condition can occur. The user should be 
aware of this latency and take this into consideration when relying on 
OCP to protect a sensitive load. The maximum period for detecting an 
OCP condition is 300 ms. 
OCP is fixed at 110% of rated current.