Calibration and Troubleshooting 
 
7-16  M370430-01  Rev B 
Calibration Procedure for Isolated Modes 
The main output and the non-isolated mode must be calibrated first. 
Isolated Voltage Monitoring Calibration 
Important: For maximum accuracy at a specific APG level, follow the 
calibration procedure again but use the intended operating voltage level in steps 
3 and 8. 
 
To calibrate the isolated voltage monitoring: 
1.  Disconnect the load and connect the voltmeter to the output. 
2.  Connect the voltmeter to the isolated voltage monitoring 
output lines (J3.5 - J3.6). 
3.  Set the Voltage APG level to 4 V. The SCPI command is: 
SYST:COMM:APR:LEV   4 
4.  Set the input source to local for voltage analog programming. 
The SCPI command is: 
SYST:REM:SOUR  LOC 
5.  Set the main output voltage to 100% by sending the 
command: 
SOUR:VOLT  <ModelVolt> 
or use the rotary adjust/Enter control. 
6.  Turn the main output to On by pressing the OUTPUT 
ENABLE Main button on the front panel or send the 
command: 
:OUTP  ON 
The OUTPUT ENABLE Main button will illuminate. 
7.  Enter the interactive calibration mode by sending the SCPI 
command: 
CAL:MON:ISOL 
8.  Press the terminal keys - or + to reduce or increase the 
monitoring value. Adjust the monitoring value displayed on 
the voltmeter that was setup in Step 1 to read as close as 
possible to 4.000 V. 
9.  Press any key other than the + or - keys to exit from 
interactive calibration mode.