Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-30 M370430-01 Rev B
Resistive-Controlled Current ISOL Setup
Activating ISOL Resistive-Controlled Current Setup
1. Turn the 9-position mode control to the CAP position or press
the rotary Adjust/Enter control if the control knob is already
at the CAP position.
is displayed on the output voltage display.
2. Turn rotary Adjust/Enter control to select (voltage
source from isolated connector).
3. Press the rotary Adjust/Enter control to commit the selected
mode. is displayed on the output voltage display.
4. Select the input source level using the rotary Adjust/Enter
control.
5. Once the desired level has been selected, press the rotary
Adjust/ Enter control to commit the change.
The display will return to the normal display.
Deactivating ISOL Current Mode
To deactivate the ISOL current mode:
1. Turn the 9-position mode control to the CAP position or press
the rotary Adjust/Enter control if the control knob is already
at the CAP position.
2. is displayed on the output voltage display.
3. Turn the rotary Adjust/Enter control to select .
4. Press the rotary Adjust/Enter control to commit the change.
The display will return to the normal display.