Calibration and Troubleshooting 
 
7-22  M370430-01  Rev B 
Calibrating the Input Voltage APG Signal 
The input voltage APG calibration is necessary to ensure accurate 
measurements when using the :MEAS : APR? and :MEAS : APR: ISOL? 
SCPI command queries to monitor the APG input signal. 
Gain Calibration 
To calibrate the VAP APG input gain: 
1.  Connect a voltmeter and a DC voltage source to the 
non-isolated voltage APG input lines (J1.9 and J1.7). 
2.  Turn the power supply On. 
3.  Set the Voltage APG level to 4 V. 
Send the SCPI command:  SYST:COMM:APR:LEV 4 
4.  Set the input source as Voltage analog APG. Send the SCPI 
command: SYST:REM:SOUR:VOLT AVOL 
5.  Set the voltage source attached to the non-isolated voltage 
APG input to 90% of the nominal voltage level, in this case 
3.6 V. 
6.  Turn the output power on. 
7.  The power supply starts up in Constant Voltage mode. 
8.  Read the input voltage value from the voltmeter connected in 
step 1. 
9.  Type the SCPI input Voltage calibration command with 
voltage noted in step 8 as the parameter “data”. 
For example, you would type : CAL: INP : ANAL 3.53 if you were 
reading 3.53 volts from the voltmeter. 
Offset Calibration 
Offset calibration of the power supply provides the best accuracy 
in low-range input voltage. 
To perform offset calibration: 
1.  Set the voltage source attached to the non-isolated Voltage 
APG input to 10% of the nominal voltage level, in this case 
0.400 V. 
2.  Turn on the output power. 
3.  Read the input voltage value from the voltmeter connected in 
step 1 of the gain calibration. 
4.  Type SCPI input Voltage calibration command with voltage 
noted in step 3 as the parameter “data”.