Remote Operation 
 
M370430-01  Rev B  5-37 
Table 5-10  Standard Event Status Register 
Set if an IEEE488.2 syntax error has been detected by 
the parser, an unrecognized header was received, or a 
group Execute Trigger was entered into the input 
buffer inside an IEEE 488.2 program message. 
 
See “Command Error List” on page B-3 for a list of 
possible error codes and descriptions. 
Reserved for possible future use by IEEE. Bit values 
are reported as zero. 
 
Operation Complete 
The Operation Complete command causes the power supply to 
generate the operation complete message in the Standard Event 
Status Register when all pending operations have been finished. 
See bit 0 in Table 5-10 for more details. 
Command: 
*OPC,    *OPC? 
Wait-to-Continue Command 
The Wait-to-Continue command prevents the power supply from 
executing any further commands or queries until the 
no-operation-pending flag is TRUE. 
Command: 
*WAI