Remote Operation
M370430-01 Rev B 5-63
5
Setting Dwell Time
The dwell time is the amount of time that is delayed between each
command during the execution of an Auto Sequence program. The
dwell time can be from 0 to 180 seconds and can be changed
during the program execution. The dwell time has a minimum step
size of 1 second.
Command:
[:]PROGram[<channel>][:STEP]:DWELl {?|<dwell time> }
Where:
<dwell time > is an integer value from 0 to 180.
For Example:
:PROG:DWEL 12 Results in 12 seconds of wait time after each
command is executed.
Storing an Auto Sequence Program:
Loading a program into memory is done by using the [: ]
PROGram [<channel>] [: RECord] : STARt command. After the
start command has been issued, the XG/XTR will record all
subsequent commands into the auto sequence program buffer. The
commands will be recorded into the buffer until the
: PROGram [<channel>] [: RECord] : STOP command is
issued.
Commands:
[:]PROGram[<channel>][:RECord]:STARt
[:]PROGram[<channel>][:RECord]:STOP
Important: If the start recording command is executed then any previously
stored Auto Sequence program is deleted.
For example:
:PROG:STAR
:VOLT 12
:OUTP ON
:PROG:DWEL 60
:OUTP OFF
:OUTP ON
:OUTP OFF