Calibration and Troubleshooting 
 
7-24  M370430-01  Rev B 
Offset Calibration 
Offset calibration of the power supply provides the best accuracy 
in low-range input current. 
To perform offset calibration: 
1.  Set the voltage source attached to the non-isolated current 
APG input to 10% of the nominal current level, in this case 
0.400 V. 
2.  Turn the output power on. 
3.  Read the input voltage value from the voltmeter connected in 
step 2 of the gain calibration. 
4.  Type SCPI input current calibration command with voltage 
noted in step 3 as the parameter “data”. 
For example, you would type   :CAL:INP:ANAL:CURR  0.43. 
After performing offset calibration, it is highly recommended that 
you repeat gain calibration. 
SCPI Command 
[:]CALibration[<channel>]:INPut:ANALog:CURRent{<voltme 
ter reading in volts>}