Local Operation 
 
3-14  M370430-01  Rev B 
Automatic Mode Crossover 
Mode crossover occurs when the unit makes the switch between 
operating modes. Crossover can occur from CV to CC or vice versa, 
depending on how the attached load resistance changes. Figure 3-3 
shows the relationships between the variables. 
For example, if the unit was initially operating in CV mode with the 
output voltage at the voltage set point and the load current below the 
I
SET
 and the attached load resistance began to decrease, the current 
would rise to meet this change in the load. The load current will increase 
in proportion to the decrease in the load resistance until the load current 
drawn reaches the current set point. At this point further decreases in 
load resistance (increased load current requirements) will cause the unit 
to cross over into CC mode where the output voltage will vary with 
changing load resistance and the load current will remain at the I
SET 
value.