Local Operation 
 
M370430-01  Rev B  3-51 
4.  Press the rotary Adjust/Enter control to commit the setting. 
The SCPI command (s) for these instructions are: 
[[:]SOURce[<channel>]]:COMBine:CSHare[:MODE] 
Setting up the Slave Units 
The output voltage and current of the slave units should be programmed 
to maximum value. During parallel operation, the slave units operate as 
a controlled current source following the controller unit's output current. 
All slave units must be configured with the SCPI command to enable 
current sharing, or use the following instructions to program front panel 
current share setup: 
1.  With the 9-position mode control at the PGM position, press 
the rotary Adjust/Enter control. 
 is displayed on the output voltage display. 
2.  Turn the Adjust/Enter conrol to select “ “ and press 
the Adjust/Enter control. 
 is displayed on the output voltage display. 
3.  Turn the rotary Adjust/Enter control to select   in the 
output current display. 
4.  Press the rotary Adjust/Enter control to commit the setting. 
The SCPI command (s) for these instructions are: 
[[:]SOURce[<channel>]]:COMBine:CSHare[:MODE] 
Important: The slave unit discussed here is not the same as the remote 
interface slave described in chapter 5. 
 
Setting Over Voltage Protection (OVP) 
The master unit's OVP should be programmed to the desired OVP level. 
The OVP of the slave units should be programmed to a higher value 
than  the  master  OVP.  When  the  controller  unit  shuts  down,  it  will 
program the outputs of the slave units to zero volts. If a slave unit shuts 
down, only that unit would shut down and the remaining slave units 
would supply all the load current if possible.