9-4
P129390-106 Operator Manual Troubleshooting
9.2.2 Too Long In
Exhaust
Occurs if chamber does not
exhaust to 0.28 bar (4 psig)
within the allotted time.
Alarm Screen
Alarm Detail
Chamber did not exhaust to
atmospheric pressure within
allotted time.
Causes and Correction:
1. Chamber drain strainer
plugged
•Clean
2. Valve malfunction
•Repair V03
3. Control out of calibration
• Recalibrate (contact
qualified Service person)
Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued)
Condition Alarm Screen on Sterilizer Control
Service Information