Troubleshooting Operator Manual P129390-106
9.2.3 Too Long In
Evacuation
Occurs if chamber does not
reach the set evacuation
level within the allotted time.
Alarm Screen
Alarm Detail
Chamber did not reach
required vacuum level within
allotted time.
Causes and Correction:
1. Water pressure less than
1.4 bar (20 Psig)
Check water supply
piping
2. Chamber drain strainer
plugged
•Clean
3. Valve malfunction
•Repair
4. Door seal not activated
Check seal
Check seal steam and
exhaust
5. Leak in plumbing
•Repair
Run a leak test
6. Control out of calibration
Recalibrate control
(contact qualified Service
person)
7. Vacuum pump
malfunction
Check vacuum pump for
proper operation
Check vacuum pump for
proper rotation
Table 9-1. Operator Troubleshooting In-Cycle Alarms (Continued)