9-9
Troubleshooting Operator Manual P129390-106
9.2.10 Steam Utility
Failure
Steam supply to sterilizer is
shut off or has failed.
Alarm Screen Check steam utility supply
and restore to proper
operation.
Causes and Corrections:
1. Steam Pressure Less than
3.4 bar (50 psig).
Check steam piping
2. Switch failure
Readjust
•Repair
9.2.11 Water Utility
Failure
Water supply to sterilizer is
shut off or has failed.
Alarm Screen Check water utility supply and
restore to proper operation.
Causes and Corrections:
1. Water pressure less than
1.4 bar (20 psig).
Check water piping
2. Switch failure
Readjust
•Repair
Table 9-1. Operator Troubleshooting In-Cycle Alarms (Continued)