6-5
Sterilizer Operation Operator Manual P129390-106
6.5 Loading Trolley
and Carrier
Instructions: Loading
1. Open sterilizer door.
2. Verify loading carrier is properly aligned upon loading trolley.
3. Align front end of loading trolley with dock at base of sterilizer.
(See Figure 6-1 and Figure 6-2).
4. Roll trolley forward until latches engage with dock.
5. Press forward Sterilizer Operationon trolley handle, loading
trolley latches to dock as carrier begins pushing into chamber.
6. Carefully push carrier into chamber.
7. Once carrier is fully inserted, turn handwheel (See Figure 6-1
and Figure 6-2) counter clockwise to lower trolley support arms
(disengaging the carrier from trolley).
8. Pull trolley away from sterilizer chamber. Loading carrier remains
in chamber.
9. Once loading trolley is fully extracted from chamber, it
disengages from dock. Move trolley away from sterilizer.
10. Sterilizer is now ready to run a cycle. Proceed to cycle
description found in S
ECTION 6.8, PREVAC CYCLE OPERATION,
PAGE 6-9.
Figure 6-1. Loading Trolley and Carrier
Handwheel
Loading Trolley
Loading Carrier
Trolley
Docking
Interface