9-11
Troubleshooting Operator Manual P129390-106
9.3 Out-of-cycle
Alarms
The following alarm screens will appear only when the sterilizer is not
processing a cycle.
Table 9-2. Operator Troubleshooting Out-of--Cycle Alarms
Condition Alarm Screen on Sterilizer Control Service Information
9.3.1 Too Long To
Close OE (or NOE)
Door
Occurs if door switch does
not make contact within
allotted time.
OE = Operating End
NOE= Non-Operating End
Alarm Screen
Alarm Detail
Door switch did not make
contact in allotted time.
Causes and Correction:
1. Door switch malfunction
Check LS1/LS3 (LS2/LS4)
Connections
Readjust LS1/LS3 (LS2/
LS4)
Repair LS1/LS3 (LS2/LS4)
2. Power door mechanism
failure
Repair mechanism
Readjust door cable
Replace motor
NOTE: Alarm may only occur
when sterilizer is hot. Test for
condition while sterilizer is both
hot and cold. Replace switch if
both 1 and 2 (above) have been
checked.