3-4
P129390-106 Operator Manual Installation Verification
3.2.3 Utility Requirements •Electric:
400 V, 3-phase, 6A / phase, 50 Hz
Water:
Pressure: 1.4 to 3.5 bar (20 to 50 psig)
Temperature: 15°C (59°F) maximum
Consumption: 49 lpm (13 gpm), peak
Steam:
Pressure: 3.5 to 5.5 bar (50 to 80 psig)
Consumption, Peak: Approximately 152 kg/hr (335 lbs/hr)
•Air:
Pressure: 5.5 to 8.3 bar (80 to 120 psig)
Consumption: 5.1 cmh (3 cfm)
3.2.4 Environmental
Conditions
Temperature: 10° to 32°C (50° to 90°F)
Humidity: 10% to 90% noncondensing
Pollution Degree: 2
Installation Category (Overvoltage Category): II
A-Weighted Sound Power Level: = 85 dBA (maximum)