6-8
P129390-106 Operator Manual Sterilizer Operation
6.7 Prevacuum
Sterilizer Cycle
Amsco Evolution Medium Steam Sterilizers are shipped with factory-
set cycle values. The cycle sequence for these cycles are given in
S
ECTION 6.8, PREVAC CYCLE OPERATION, PAGE 6-9. Refer to Table 6-4,
below, for factory set, qualified cycle settings.
The sterilization cycles listed in Table 6-4 have been validated using
techniques documented in EN 285:2006. If different cycle
parameters (sterilize time and dry time only) other than those in
Table 6-4 are required, it is the responsibility of the healthcare
facility to validate the cycle. Reference EN ISO 17665-1:2006 for a
guide to validating sterilization cycles and to ensure drying
efficiencies are met.
NOTE: Contact STERIS for information on a wide range of
education/training programs designed to meet the educational needs
of healthcare facilities.
Table 6-4. Prevacuum Sterilizer Cycles and Cycle Values
Cycle Sterilize Temp. Sterilize Time Dry Time Validation Standard
PREVAC 134°C (273°F) 3 Min., 30 Sec. 30 Min. or 20 Min. EN 285:2006
PREVAC 121°C (250°F) 30 Min. 30 Min. or 20 Min. EN 285:2006
LIQUID 121°C (250°F) 45 Min. N/A N/A