Sterilizer Operation Operator Manual P129390-106
6.6 Loading Trolley
and Carrier
Instructions: Unloading
1. Open chamber door.
2. Move loading trolley forward until it engages with dock.
3. Once loading trolley engages with dock, loading trolley latches
to dock as it begins pushing into chamber.
4. Push trolley fully into chamber
5. Turn handwheel clockwise until loading carrier is fully lifted from
supports in chamber.
6. Withdraw carrier and trolley from chamber.
7. Loading trolley disengages from dock when fully withdrawn from
chamber. Pull trolley away from sterilizer.
8. Close chamber door.
9. Transfer load from sterilizer area.
WARNING – BURN HAZARD:
• Sterilizer, shelves and loading
carrier will be hot after cycle is
run. Always wear protective
gloves and apron when
removing a processed load.
Protective gloves and apron
must be worn when reloading
sterilizer following the
previous operation.
• Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door is
opened to minimize contact
with steam vapor.
WARNING – SLIPPING
HAZARD: To reduce slippery
floor conditions, keep floors
dry by immediately wiping up
any spilled liquids or
condensation in sterilizer
loading or unloading area.