7-4
P129390-106 Operator Manual Cycle and Control Value Programming
7.3 Change Cycle
Values
Important: Applicable cycles have been validated to satisfy the
requirements outlined in Table 4-1. If cycle parameters (sterilize time
or dry time) other than those in Table 7-1 are required, it is the
responsibility of the healthcare facility to validate the cycle.
Reference EN ISO 17665-1:2006 standards for a guide to validating
sterilization cycles and to ensure drying efficiencies are met.
7.3.1 Overview Refer to screens shown in SECTION 7.3.2, CHANGE CYCLE VALUES
PROCEDURE, PAGE 7-5 for more detail.
Press anywhere on the touch screen pad if the sterilizer is in
Standby. The control advances to the ready. At screen #1, press the
Options touch screen pad (upper right corner), and the control
advances to Machine Setup screen; access the cycle values by
pressing Change Cycle Values. Press one of the displayed cycles
(or Next to find a cycle not currently on the display). Once the cycle
has been selected, the screen changes to show basic cycle
information. Select the value to change:
• Cycle Name (name of the cycle),
• Sterilize Temp (sterilize exposure temperature),
• Sterilize Time (sterilize exposure time) or
• Dry (drying phase time).
Press the touch screen pad; the display changes to a screen for
making these changes. Once all selections have been entered,
press the OK touch screen pad to accept or Exit to abandon
changes and return to change cycle values screen.