7-2
P129390-106 Operator Manual Cycle and Control Value Programming
7.1 Cycle Values Amsco Evolution sterilizers are shipped with the default cycle values
shown in Table 7-1.
The Prevacuum Configuration Amsco Evolution sterilizer is equipped
with the following factory programmed sterilization cycle values
(Table 7-1).
1.
Cross reference model number to chamber size and capacity using table below:
Table 7-1. Prevacuum Sterilizer Cycle Values
Cycle Sterilize Temp. Sterilize Time Dry Time
Validation
Standard
PREVAC 134°C (273°F) 3 Min., 30 Sec 30 Min. or 20 Min. EN 285:2006
PREVAC 121°C (250°F) 30 Min. 30 Min. or 20 Min. EN 285:2006
LIQUID 121°C (250°F) 45 Min. N/A N/A
Table 7-2. Test Cycles
Cycle Sterilize Temp. Sterilize Time Dry Time
Validation
Standard
Leak Test
a
N/A N/A N/A EN 285:2006
Bowie-Dick Test
a
134°C (273°F) 3 Min., 30 Sec. 1 Min. EN 285:2006
Bowie-Dick Test
Warm-up
a
134°C (273°F) 3 Min., 30 Sec. 1 Min. N/A
a. Not adjustable.
Table 7-3. Sterilizer Chamber Volume
Model Chamber Size
Number of
Sterilization Modules
HC-600
660 x 660 x 991 mm (26 x 26 x 39") 6
HC-800
660 x 660 x 1245 mm (26 x 26 x 49") 8
HC-1000
660 x 660 x 1549 mm (26 x 26 x 61") 10
HC-900 660 x 950 x 1067 mm (26 x 37.5 x 42") 9
HC-1200 660 x 950 x 1372 mm (26 x 37.5 x 54") 12
HC-1500 660 x 950 x 1676 mm (26 x 37.5 x 66") 15