7-3
Cycle and Control Value Programming Operator Manual P129390-106
7.2 Change Values Important: Applicable cycles have been validated to satisfy the
requirements outlined in Table 4-1 (page 4-2). If cycle parameters
(sterilize time or dry time) other than those in Table 7-1 are required,
it is the responsibility of the healthcare facility to validate the cycle.
Reference EN ISO 17665-1:2006 for a guide to validating sterilization
cycles and to ensure drying efficiencies are met.
The Change Values procedure described in the following pages can
be used to adjust cycle values within a limited range and to make
selections affecting the operating style of the unit. Refer to Table 7-1
for cycle values.
The Change Values Procedure can be used to adjust the following
values:
Sterilization Exposure Temperature
Important: Sterilization temperature is limited to certain values, refer
to S
ECTION 7.3, CHANGE CYCLE VALUES, PAGE 7-4.
Sterilization Exposure Time
Important: Sterilize time is limited to a predetermined range of
values, refer to S
ECTION 7.3, CHANGE CYCLE VALUES, PAGE 7-4.
Dry Time
Important: Dry time is limited to a predetermined range of values,
refer to S
ECTION 7.3, CHANGE CYCLE VALUES, PAGE 7-4.
Other Information
• Change Values can also be used to make adjustments to the
Machine Setup values shown on Table 7-4.
• Default cycle values are shown on Table 7-1.
• Change Values screens are accessed by pressing Change
Values button on Options screen.