9-21
Troubleshooting Operator Manual P129390-106
9.4.10 OE/NOE
Door Switches
Malfunction
Occurs if control senses
OE or NOE door is in the
open and closed position
simultaneously.
Alarm Screen
Alarm Detail
Door switches do not agree.
Causes and Correction:
1. Door switch malfunction
Check LS1(LS3) connections
for OE door or LS2 (LS4) For
NOE Door
Readjust door switches
Determine which switch is in
the incorrect state, and
replace that switch.
9.4.11 Exhaust
Rate Too Slow
Occurs if liquid cycle slow
exhaust rate is too slow.
Alarm Screen
Alarm Detail
Chamber exhausted slower than
the expected rate.
Causes and Correction:
1. Valve malfunction Repair
V40
2. Valve malfunction Repair
V03
3. Chamber steam trap
malfunction Repair
4. Control out of calibration
Recalibrate (contact qualified
service person)
Table 9-3. Operator Troubleshooting Sensor Alarms (Continued)
Condition Alarm Screen on Sterilizer Control Service Information