P129390-106 Operator Manual Sterilizer Operation
7.
Cross reference model number to chamber size and capacity using table below:
Table 6-1. Prevacuum Sterilizer Cycle Values
Cycle Sterilize Temp. Sterilize Time Dry Time
Validation
Standard
PREVAC 134°C (273°F) 3 Min., 30 Sec 30 Min. or 20 Min EN 285:2006
PREVAC 121°C (250°F) 30 Min. 30 Min. or 20 Min EN 285:2006
LIQUID 121°C (250°F) 45 Min. N/A N/A
Table 6-2. Test Cycles
Cycle Sterilize Temp. Sterilize Time Dry Time
Validation
Standard
Leak Test
a
N/A N/A N/A EN 285:2006
Bowie-Dick Test
a
134°C (273°F) 3 Min. 30 Sec. 1 Min. EN 285:2006
Bowie-Dick Test
Warm-up
a
134°C (273°F) 3 Min. 30 Sec. 1 Min. N/A
a. Not adjustable.
Table 6-3. Sterilizer Chamber Volume
Model Chamber Size
Number of
Sterilization Modules
HC-600 660 x 660 x 991 mm (26 x 26 x 39") 6
HC-800 660 x 660 x 1245 mm (26 x 26 x 49") 8
HC-1000 660 x 660 x 1549 mm (26 x 26 x 61") 10
HC-900 660 x 950 x 1067 mm (26 x 37.5 x 42") 9
HC-1200 660 x 950 x 1372 mm (26 x 37.5 x 54") 12
HC-1500 660 x 950 x 1676 mm (26 x 37.5 x 66") 15