EasyManua.ls Logo

Amsco Evolution HC-600 - Electrical Service; Sterilizer Final Check; Cycle Operation

Amsco Evolution HC-600
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-2
P129390-106 Operator Manual Installation Verification
3.1.3 Electrical Service The protective earth ground must be connected to terminal block
TB-1 in the sterilizer power box.
Three-phase power for the sterilizer must meet specifications on
the equipment drawing.
Verify proper rotation of the vacuum pump by observing pump
rotor shaft.
Important: Three-phase service requires a clearly marked
disconnect with lockout/tagout capability located near the
sterilizer with easy access for maintenance personnel. If more
than one sterilizer is connected to the same supply, each should
be equipped with independent lock/tagout capability,
3.1.4 Sterilizer Final Check Chamber leveled properly.
Door opens and closes smoothly.
Door locked switches adjusted correctly.
Chamber strainer in place.
Rack and shelves and/or loading carrier operate correctly.
Paper loaded in printer.
Printer ribbon properly installed.
Warranty labels properly applied.
Verify programming and set up support proper configuration of
unit.
3.1.5 Cycle Operation Unit powers up correctly.
Run Leak Test cycle leak rate is to be less than 1.3 mbar/min
(0.4 mm Hg/minute).
Verify operation of a typical cycle (134° C [273°F] prevacuum).

Table of Contents

Related product manuals