9-19
Troubleshooting Operator Manual P129390-106
9.4.6 Chamber
Temperature Probe
Failure
Occurs if chamber
temperature reading
outside the normal range.
Alarm Screen
Alarm Detail
RTD probe, RTD1, output is
outside normal range.
Causes and Correction:
1. Loose connection in probe
wiring
• Repair
2. Probe failed
• Replace
• Recalibrate
3. Control out of calibration
• Recalibrate (contact qualified
Service person)
9.4.7 Jacket
Temperature Probe
Failure
Occurs if jacket
temperature reading is
outside the normal range.
Alarm Screen
Alarm Detail
RTD probe, RTD3, output is
outside normal range.
Causes and Correction:
1. Loose connection in probe
wiring
• Repair
2. Probe failed
• Replace
• Recalibrate
3. Control out of calibration
• Recalibrate (contact qualified
Service person)
Table 9-3. Operator Troubleshooting – Sensor Alarms (Continued)
Condition Alarm Screen on Sterilizer Control Service Information