6-10
P129390-106 Operator Manual Sterilizer Operation
NOTE: Countdown timer on the display is estimated; timer self-
corrects estimated time at the beginning of each phase.
VACUUM/PRESSURE PULSES #1 - #7 – Three vacuum pulses
(drawing from 0.1 bar to 0.8 bar) are followed by four pressure
pulses (0.1 bar to 1.1 bar).
CHARGE – Chamber is charged with steam. Start of steam
charge is printed.
STERILIZE – Start of sterilize exposure is printed when the
chamber reaches sterilization temperature. Chamber
temperature is printed every minute.
FAST EXHAUST – Start of exhaust is printed and chamber is
exhausted to 0.28 bar (4 psig).