6-10
P129390-106 Operator Manual Sterilizer Operation
NOTE: Countdown timer on the display is estimated; timer self-
corrects estimated time at the beginning of each phase.
VACUUM/PRESSURE PULSES #1 - #7 Three vacuum pulses
(drawing from 0.1 bar to 0.8 bar) are followed by four pressure
pulses (0.1 bar to 1.1 bar).
CHARGE Chamber is charged with steam. Start of steam
charge is printed.
STERILIZE Start of sterilize exposure is printed when the
chamber reaches sterilization temperature. Chamber
temperature is printed every minute.
FAST EXHAUST Start of exhaust is printed and chamber is
exhausted to 0.28 bar (4 psig).