After switching an application, Answer incoming ARP requestsAnswer incoming ARP requests in the Settings screen is
selected.
After switching to the RFC 2544 and SAT(Y1564) application, the following settings are
selected.
Accept Network Master configuration framesAccept Network Master configuration frames in the Settings screen
Use local source addresses for destination on remote sideUse local source addresses for destination on remote side in the Advanced
screen
After switching to the Reflector application, the following settings in the Swap screen
are selected.
Swap all MAC addressesSwap all MAC addresses
Swap IP addressesSwap IP addresses
Swap UDP/TCP portsSwap UDP/TCP ports
For other settings, the same values are used after switching the application.
66..1313..44 ProcedureProcedure
This section explains the procedure to control the Network Master through a
network using Discovery application. In the example here, Network Master B
discovers Network Master A. Network Master B switches the application of
Network Master A to RFC2544 and performs RFC2544 test.
Network Master A SetupNetwork Master A Setup
1. Turn on Network Master A.
2. Start one of Ethernet applications on Network Master A.
3. Set up interface on the PortPort screen and establish Ethernet link.
4. Touch the SettingsSettings on the SETUP screen.
5. Touch the In-band SetupIn-band Setup tab.
6. Select Allow in-band controlAllow in-band control.