7.3 64G DEMUX Commands
7-21
7
Remote Commands
:DEMux:DATA:TERMination <term>
Parameter <term>=<CHARACTER PROGRAM DATA>
GND 50 to GND
VARiable 50 to Variable Voltage
Function Sets the data input termination condition.
Example To set the termination condition of Data Interface to GND:
> :DEMux:DATA:TERMination GND
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:TERMination?
Response <term>=<CHARACTER RESPONSE DATA>
GND, VAR
Function Queries the data input termination condition.
Example To query the termination condition of Data Interface:
> :DEMux:DATA:TERMination?
< GND
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:TLEVel <numeric>
Parameter <numeric>=<DECIMAL NUMERIC PROGRAM DATA>
–2.50 to 3.50 –2.50 to 3.50 V/0.01 V Step
Function Sets the termination voltage when the input termination condition is set to
Variable.
Example To set the termination voltage to –2.5 when the input termination condition
is set to Variable:
> :DEMux:DATA:TLEVel -2.5
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:TLEVel?
Response <numeric>=<NR2 NUMERIC RESPONSE DATA>
Function Queries the termination voltage when the input termination condition is set
to Variable.
Example To query the termination voltage when the input termination condition is
set to Variable:
> :DEMux:DATA:TLEVel?
< -2.50
Compatibility Compatible with MU182040A and MU182041A.