Chapter 7 Remote Commands
7-30
:DEMux:MEASure:MREStart <data>,<clock>
Parameter <data>=< BOOLEAN PROGRAM DATA >
0 or OFF Set measurement restart at Data Threshold change to
OFF.
1 or ON Set measurement restart at Data Threshold change to
ON.
<clock>=< BOOLEAN PROGRAM DATA >
0 or OFF Set measurement restart at Clock Delay change to OFF.
1 or ON Set measurement restart at Clock Delay change to ON.
Function Sets measurement restart conditions.
Example To set Data Interface measurement restart conditions:
> :DEMux:MEASure:MREStart 0,0
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:MEASure:MREStart?
Response <data>=< NR1 NUMERIC RESPONSE DATA >
<clock>=< NR1 NUMERIC RESPONSE DATA >
0 Measurement restart OFF
1 Measurement restart ON
Function Queries measurement restart conditions.
Example To query Data Interface measurement restart conditions:
> :DEMux:MEASure:MREStart?
< 0,0
Compatibility Compatible with MU182040A and MU182041A.