7.4 Automatic Measurement Commands
7-97
7
Remote Commands
:SENSe:MEASure:BATHtub:CURRent?
Response <boolean>=<NR1 NUMERIC RESPONSE DATA>
0, 1
Function Queries the graph display update period during Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:CURRent?
< 1
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:JCALculation <setting>
Parameter <setting>=<DECIMAL NUMERIC PROGRAM DATA>
7 to 20 E–7 to E–20
Specified as E−n where n: 7 to 20/1 step
Function Sets the error rate used to calculate jitter at Bathtub measurement.
Example of use To set jitter calculation error rate to E–14:
> :SENSe:MEASure:BATHtub:JCALculation 14
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:JCALculation?
Response <setting>=<NR1 NUMERIC RESPONSE DATA>
7 to 20
Function Queries the error rate used to calculate jitter at Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:JCALculation?
< 14
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:J2Meas <mode>
Parameter <mode>=<CHARACTER PROGRAM DATA>
ACTual Calculates J2 from BER measurement point.
ESTimate Calculates J2 from Best Fit Line.
Function Selects J2 calculation method.
Example To set J2 calculation method to Actual:
> :SENSe:MEASure:BATHtub:J2Meas ACTual
Compatibility Incompatible with existing models.