3-7 Input Related Spurious (IRS) Signals Spectrum Analyzer Performance Verification
3-14 PN: 10580-00303 Rev. D Model MS20xxB MM
24. Press the Marker main menu button and then the Peak Search submenu button.
25. Record the Marker amplitude reading of 2072.5 MHz in Table A-15 on page A-6 for MS2034B or
Table A-40 on page A-19 for MS2035B in Appendix A.
26. Press the Freq main menu button and then the Start Freq submenu button.
27. Use the numeric keypad to enter 1222.5 and then press the MHz submenu button.
28. Press the Stop Freq submenu button.
29. Use the numeric keypad to enter 2072 and then press the MHz submenu button.
30. Wait until one sweep is completed.
31. Press the Marker main menu button and then the Peak Search submenu button.
32. Record the Marker amplitude reading into the Measured Value column for 1222.5 MHz to 2072 MHz in
Appendix A.
• MS2034B – Table A-15, ā€œSpectrum Analyzer Input Related Spurious (IRS) Signalsā€ on page A-6
• MS2035B – Table A-40, ā€œSpectrum Analyzer Input Related Spurious (IRS) Signalsā€ on page A-19
33. Calculated the input related spurious level using the formula below:
Input Related Spurious dBc = Marker 1 Reading – Amplitude Reading of 2072.5 MHz
34. Record the calculated value into the Calculated IRS column for 1222.5 MHz to 2072 MHz in Table A-15
for MS2034B or Table A-40 for MS2035B in Appendix A and verify that it is <
–60 dBc.
35. Repeat Step 26 through Step 34 setting a start frequency of 2073 MHz and a stop frequency of
2922.5 MHz and record the results into the appropriate columns for 2073 MHz to 2922.5 MHz in
Table A-15 for MS2034B or Table A-40 for MS2035B in Appendix A.
1220 MHz Input Related Spurious Check
36. Set the MG3692X frequency to 1500 MHz CW.
37. Press the Freq main menu button and then the Center Freq submenu button.
38. Use the numeric keypad to enter 1500 and then press the MHz submenu button.
39. Press the Span submenu button.
40. Use the numeric keypad to enter 2 and then press the MHz submenu button.
41. Press the BW main menu button and then the RBW submenu button.
42. Use the numeric keypad to enter 3 and then press the kHz submenu button.
43. Press the VBW submenu button.
44. Use the numeric keypad to enter 100 and then press the Hz submenu button.
45. Press the Shift key, press the Trace (5) key and then press the Trace A Operations submenu button.
46. Press the # of Averages submenu button.
47. Use the numeric keypad to enter 5, and then press the Enter key.
48. Wait until the Trace Count displays ā€œ5/5ā€.
49. Press the Marker main menu button and then Peak Search submenu button.
50. Record the amplitude at 1500 MHz in Table A-15 for MS2034B or Table A-40 for MS2035B in
Appendix A.
51. Press the
Freq main menu button and then the Start Freq submenu button.
52. Use the numeric keypad to enter 1219 and then press the MHz submenu button.
53. Press the Stop Freq submenu button.
54. Use the numeric keypad to enter 1221 and then press the MHz submenu button.

Table of Contents

Other manuals for Anritsu VNA Master MS2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu VNA Master MS2024B and is the answer not in the manual?

Anritsu VNA Master MS2024B Specifications

General IconGeneral
BrandAnritsu
ModelVNA Master MS2024B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals