3-7 Input Related Spurious (IRS) Signals Spectrum Analyzer Performance Verification
3-16 PN: 10580-00303 Rev. D Model MS20xxB MM
87. Press the Marker main menu button and then Peak Search submenu button.
88. Record the Marker amplitude reading into the Measured Value column for 2821 MHz to 2823 MHz in
Table A-15 on page A-6 for MS2034B or Table A-40 on page A-19 for MS2035B in Appendix A.
89. Calculated the input related spurious level using the formula below:
Input Related Spurious dBc = Marker 1 Reading Amplitude Reading of 1500 MHz
90. Record the calculated value into the Calculated IRS column for 2821 MHz to 2823 MHz in Table A-15 for
MS2034B or Table A-40 for MS2035B in Appendix A and verify that it is <
–52 dBc.
2506 MHz Input Related Spurious Check
91. Set the MG3692X frequency to 2421.5 MHz CW.
92. On the VNA Master, press the Freq main menu button and then the Center Freq submenu button.
93. Use the numeric keypad to enter 2421.5 and then press the MHz submenu button.
94. Press the Span submenu button.
95. Use the numeric keypad to enter 2 and then press the MHz submenu button.
96. Press the BW main menu button and then the RBW submenu button.
97. Use the numeric keypad to enter 10 and then press the kHz submenu button.
98. Press the VBW submenu button.
99. Use the numeric keypad to enter 1 and then press the kHz submenu button.
100. Press the Shift key, press the Trace (5) key and then press the Trace A Operations submenu button.
101. Press the # of Averages submenu button.
102. Use the numeric keypad to enter 5, and then press the Enter key.
103. Wait until the Trace Count displays “5/5”.
104. Press the Marker main menu button and then the Peak Search submenu button.
105. Record the Marker amplitude reading of 2421.5 MHz in Table A-15 for MS2034B or Table A-40 for
MS2035B in Appendix A.
106. Press the Freq main menu button and then the Start Freq submenu button.
107. Use the numeric keypad to enter 2505 and then press the MHz submenu button.
108. Press the Stop Freq submenu button.
109.
Use the numeric keypad to enter 2507 and then press the MHz submenu button.
110. Press the Shift key, press the Trace (5) key and then press the Trace A Operations submenu button.
111. Press the # of Averages submenu button.
112. Use the numeric keypad to enter 5, and then press the Enter key.
113. Wait until the Trace Count displays “5/5”.
114. Press the Marker main menu button and then the Peak Search submenu button.
115. Record the Marker amplitude reading into the Measured Value column for 2505 MHz to 2507 MHz in
Table A-15 for MS2034B or Table A-40 for MS2035B in Appendix A.
116. Calculated the input related spurious level using the formula below:
Input Related Spurious dBc = Marker 1 Reading Amplitude Reading of 2421.5 MHz
117. Record the calculated value into the Calculated IRS column for 2505 MHz to 2507 MHz in Table A-15 for
MS2034B or Table A-40 for MS2035B in Appendix A and verify that it is <
–39 dBc.

Table of Contents

Other manuals for Anritsu VNA Master MS2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu VNA Master MS2024B and is the answer not in the manual?

Anritsu VNA Master MS2024B Specifications

General IconGeneral
BrandAnritsu
ModelVNA Master MS2024B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals