25
Line, Sub-line Parameter Setting,
basic
Setting,
range
Authorization
32 04 Limit for inspection
requests
9999 0-9999 C
32 05 Counter for inspection
request
- 0-9999 -
Line 33: Dose size
33 01 Dosing time, descaler 6 0-99 seconds BC
33 02 Dosing time rinse-aid 4 0-99 seconds BC
33 03 Dosing time, process agent 30 0-99 seconds BC
33 04 Dosing time spray agent 0.7 0.0-9.9 seconds BC
Line 34: Empty container alarm
34 01 Alarm descaler 0125 00 = no alarm,
01XX* = warning,
02 = stop
C
;;±LVWKHQXPEHURISURFHVVHVWKDWFDQEHUXQDIWHUDQDODUP
34 02 Alarm chemical agent 0200 00 = no alarm,
01XX* = warning,
02 = stop
C
34 03 Alarm rinse-aid 0125 00 = no alarm,
01XX* = warning,
02 = stop
C
34 04 Alarm, process agent 0110 00 = no alarm,
01XX* = warning,
02 = stop
C
34 05 Alarm spray agent 0110 00 = no alarm,
01XX* = warning,
02 = stop
C
;;±LVWKHQXPEHURISURFHVVHVWKDWFDQEHUXQDIWHUDQDODUP
Lines 35-49: Disinfection
35 - Normal disinfection method For future use
36 - Reserve disinfection method For future use
37 - Viewing mode 7 1-7 C
38 - Lower disinfection
temperature
91 80-95°C C
39 - Upper disinfection
temperature
93 85-97°C C
40 - Disinfection temperature 91 80-95°C C
41 - Disinfection time 35 0-1,800 seconds C
42 - Max preheat time, steam
generator
50 0-99 seconds C
43 - Power sharing, steam
generator
1 0 = Not active,
1 = Active
C
44 1 Cooling method 2 0 = None,
1 = Internal,
2 = External,
3 = Extended
C