26
Line, Sub-line Parameter Setting,
basic
Setting,
range
Authorization
44 2 Global external cooling time 120 0-180 sec C
44 3 Global extended cooling
(after process)
120 0-180 sec C
45 1 Handling auto-disinfection 0 0 = None,
1 = Warning,
2 = Warning +
autostart
C
45 2 Auto-disinfection permitted
until
8 Time 0-23 C
45 3 Auto-disinfection permitted
until
20 Time 0-23
46 - Disinfection to A0 0 0, 60, 600 C
48 - Dosage, chemical agent 70 0-99 seconds C
49 - Chemical action time 0 0-600 seconds C
Lines 50-57: Tank and water
50 - Tank level at rest 0 0 = empty tank,
1 = medium level,
2 = high level
C
51 - :DWHUW\SH¿OOLQJDWUHVW 1 1 = cold, 2 = hot,
3 = cold and hot
C
52 - Mix ratio 50 0-100% cold water C
53 01 7\SHRIULPÀXVKLQJ 1 0 = None,
1 = Internal
C
54 - Addition of softened water 0 0 = none, 1 = in
tank, 2 = in steam
generator
C
55 - Time for soft water 5 0-99 seconds C
56 - Flushing time cooling 0.0 0.0-99.9 seconds C
57 01 Flushing time in the
LQVWUXFWLRQ³3XOVHÀXVKLQJ
to level”
For future use
57 02 Pause time in indication
³3XOVHÀXVKLQJWROHYHO´
For future use
57 03 Pulses in the instruction
³3XOVHÀXVKLQJWROHYHO´
For future use
Lines 60-63: Error statistics
60 - Time of error 0-9999 seconds -
61 - Instruction for last error 0 = Not in
ÀXVKLQJSURJUDP
1-9999 = program
line
-
62 - Counter, total number of
errors
0-9999 -
63 01 Counter for error code F1 0-99 errors C