Maintenance
7-1
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
7 Maintenance
Cleaning the sensor
Maintenance of the HYGROPHIL
®
F 5673 is limited to cleaning the moisture-
sensitive layer on the tip of the sensor L 166x.
The cleaning intervals depend on how dirty the sample gas is.
Handle the sensor very carefully. The moisture-sensitive layer is very sus-
ceptible to knocks and scratches. Never use pointed or sharp edged objects
to clean the sensor.
Damage may result in incorrect measurements.
Take the probe out of the measuring line.
(1)
Use soft cleaning cloths only.
Clean the sensor tip with a soft cloth twisted to a tip, preferably one
impregnated with alcohol (isopropyl).
BARTEC BENKE recommends OpticPads CT811 supplied by CleanTex,
or similar products
Then wipe the surface several times with a soft cloth to remove any
residues of alcohol.
Use the cloths only once!
Reinsert the sensor in the measuring line.
(1)
(1)
When removing and reinserting the sensor, follow the instructions provid-
ed with the accessories used (e.g. sensor retraction tool).
Moisture-sensitive layer