Programming
5-17
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
5.6.3.2 Pressure
Define whether to use a pressure sensor or not.
Under Default value, enter a fixed value. This fixed value will be applied if
no sensor is used.
Define the measurement range for the current input of pressure.
The minimum value corresponds to an input current of 0 and 4 mA
respectively, the maximum value corresponds to 20 mA.
5.6.3.3 CO
2
content
Define whether to use a CO
2
sensor or not.
Under Default value, enter a fixed value. This fixed value will be applied if
no sensor is used.
Define the measurement range for the current input of CO
2
content.
The minimum value corresponds to an input current of 0 and 4 mA
respectively, the maximum value corresponds to 20 mA.
Default values
The fixed value that is entered for a measuring variable will be applied if no
sensor is used.
If a sensor should be used but is not connected or if a connected sensor is
defective, the fixed value will automatically be used. If a fixed value is used,
an asterisk appears behind the measured value in the display
(see section 4.3.1.4).