Programming
5-17
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
5.6.3.2 Pressure
● Define whether to use a pressure sensor or not.
● Under Default value, enter a fixed value. This fixed value will be applied if
no sensor is used.
● Define the measurement range for the current input of pressure.
The minimum value corresponds to an input current of 0 and 4 mA
respectively, the maximum value corresponds to 20 mA.
5.6.3.3 CO
2
content
● Define whether to use a CO
2
sensor or not.
● Under Default value, enter a fixed value. This fixed value will be applied if
no sensor is used.
● Define the measurement range for the current input of CO
2
content.
The minimum value corresponds to an input current of 0 and 4 mA
respectively, the maximum value corresponds to 20 mA.
Default values
The fixed value that is entered for a measuring variable will be applied if no
sensor is used.
If a sensor should be used but is not connected or if a connected sensor is
defective, the fixed value will automatically be used. If a fixed value is used,
an asterisk appears behind the measured value in the display
(see section 4.3.1.4).