EasyManua.ls Logo

Basler BE1-11g - Page 429

Basler BE1-11g
672 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9424200994 Rev N 417
Figure 289. BESTlogicPlus Settings (Vx Fundamental Mode)
Step 3: Use BESTCOMSPlus to open the Protection, Voltage, Overvoltage (59X-1) screen and send
the first row of test settings in Table 167 to the BE1-11g.
Table 167. Pickup Test Settings (Vx Fundamental Mode)
Pickup Setting
Time Delay
125 V
50 ms
75 V
50 ms
30 V
50 ms
Step 4: Prepare to monitor the 59X-1 function operation. Operation can be verified by monitoring OUT2
(see Figure 289).
Step 5: Connect and apply a 120 Vac, single-phase voltage source to the Vx input, terminals C17
(polarity) and C18 (non-polarity).
Step 6: Slowly increase the voltage until OUT2 closes and record the pickup. Verify that there is a 59X-
1-AUX target on the front-panel display. Slowly decrease the voltage until OUT2 opens and
record the dropout. Reset the target.
Step 7: Verify the pickup and dropout accuracy at 70 Vac for a pickup setting of 75 V and 25 Vac for a
pickup setting of 30 V as listed in Table 167. Record the results.
Step 8: (Optional.) Repeat steps 1 through 7 for the B-phase and C-phase voltage inputs.
Step 9: (Optional.) Repeat steps 1 through 8 for settings group 1, 2, and 3.
Step 10: (Optional.) Repeat steps 1 through 9 for 59X-2, 59X-3, and 59X-4.
Timing Verification (Vx Fundamental Mode)
Step 1: Use BESTCOMSPlus to open the Protection, Voltage, Overvoltage (59X-1) screen and send
the first row of test settings in Table 168 to the BE1-11g for settings group 0.
Table 168. Timing Test Settings (Vx Fundamental Mode)
Pickup Setting
Time Delay
125 V
2,000 ms
125 V
5,000 ms
125 V
10,000 ms
Step 2: Prepare to monitor the 59X-1 timings. Timing accuracy is verified by measuring the elapsed
time between a sensing voltage change and OUT1 closing.
Step 3: Connect and apply a 120 Vac, single-phase voltage source to terminals C17 (polarity) and C18
(non-polarity).
Step 4: Step the voltage up to 130 volts. Measure the time delay and record the result.
BE1-11g Auxiliary Overvoltage (59X) Test

Table of Contents

Related product manuals