07-19-B1-729 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL C-1
Appendix C
Configuration Options Data SheetAppendix C
Use the data sheet pages herein to record the Configuraion Option settings
described in Chapter 5, “Configuration Option Feature”.
Note: Not all of the options below are available in all software versions. Factory
defaults may not match pump defaults.
Serial Number: Date:
Configuration Options Factory Default Values Current Settings
Occlusion Alarm Level Level 1
R Level 1 R Level 2 R Level 3
Audible Switchover Off
R Off R On
Number of Automatic Restarts 3 Range 0-9, record value:
Door Open Required Off
R Off R On
Air Bubble Alarm Size Norm
R Norm (avg. 75 μL) R Min (avg. 50 μL)
Alarm Off Interval 1 Range 1-7, record value:
Alert Off Interval 7 Range 1-7, record value:
Maximum Rate of Infusion 1999 mL/hr Range 1- 1999 mL/hr, record value:
Maximum VTBI 9999 mL Range 1- 9999 mL, record value:
Flow Check Display Off
R Off R On
Baud Rate 9600
R 300 R 1200 R 2400 R 4800 R 9600
Computer Control Disabled
R Disabled R Off w/Alarm R On w/Alarm