LB 134 7. Probes for the LB 134
95
Figure 7.10: Dose rate probe LB 1236-H10
The detector background is determined at the factory for each
probe and can be taken from the supplied data sheet. The de-
tector background values are entered in the Background sub-
menu.
Connect the detector with the 8-pin connector Fischer to the
turned off LB 134. After power on of the LB 134, the connected
detector is checked and identified by the software, and the re-
spective parameters are loaded.
If there is a detector failure (defect), the entire probe has to be
replaced. The counter tubes cannot be replaced by the custom-
er, but only by BERTHOLD service technicians.
Technical data of the probe LB 1236-H10
Calibration factor Intrinsic background
μSv/h/cps cps
0.214 < 0,1 cps
Nominal operating ranges probe LB 1236-H10
Parameter
Range
Max.
Devia-
tion
Energy of gammas
30 keV 1,3 MeV
± 30%
Preferred Direction
Vertical to counter axis
Direction of inco-
ming radiation
± 45° to main radiation
direction, blue ring around
probe
± 20%
Ambient Tempera-
ture
-10 to +60° C
± 20%