3.5 Possible Problems during Commissioning
The following errors or error messages may be encountered during commissioning:
o For scint detectors: If detector foil is defective and
light enters, the high voltage turns off, count rate
is zero. Then the device must be turned off.
o For gas-filled contamination detectors: If detector
is OK, the detector foil must be smooth, tense
and plane. A slack, wrinkly foil indicates that the
counter tube is
damaged. Please call Service!
Possibly, HV unit of detector is defective. Replace the de-
tector if the HV unit is defective.
The set background is too high. To check this, push the
Info button to view the gross data.
Connection electronics <--> check probe
Count rate in alpha channel > 750 cps. Above this count
rate, no meaningful measurement is possible in the beta
channel.