EasyManua.ls Logo

Berthold LB 134 - Exceeding of Limit Values

Berthold LB 134
157 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LB 134 5. Contamination Measurements
50
a) whether the same radionuclide has been set, and
b) whether the source used for calibration meets the above
mentioned requirements.
When measuring area contaminations, the extent of the con-
tamination relative to the standardized area of the calibration
source of 100 cm² (in the setting A-100) also plays a role.
For measurements according to §44 of the Radiation Protection
Ordinance (Germany), the averaging area can be up to
300 cm². The measurement area of the probe LB 1342 is 118
mm x 145 mm (171 cm²) at a transmission of 80%. In the
LB 1343 probe the active measurement area is 345 cm².
For contaminations less than 100 cm² this averaging is done
automatically due to the aforementioned calibration conditions.
A contamination of, say, 10 Bq/cm², but an expansion of only 50
cm², is displayed as 5 Bq/cm² (distributed over 100 cm²) as a
result of this averaging.
However, if the extent of the contamination is greater than 100
cm², the monitor overevaluates: It shows at 10 Bq/cm², spread
over 150 cm², for example, 15 Bq/cm².
5.4 Exceeding of Limit Values
Any exceeding of user-defined limit values is indicated on the
display by a flashing measured value ( or ). Moreover, an
alarm signal is output (visual as flashing LED, audible as short
alarm beep), if preset (see System menu: Parameter->Alarm).
The alarm is displayed in any case, no matter which measure-
ment mode you are in.
Figure 4.17: Exceeding of limit values
If the entrance window is defective (light enters the counting
chamber) and the count rate is exceeded for a short time, the
scintillation detector is turned off, i.e. the high voltage is turned
off. In this case, the entrance window must be replaced.

Table of Contents

Related product manuals