Date/Time: (6 bytes)
(Memory number: is sent with F
2
C)
Cycle number: 1...2000 (ushort) (2 bytes)
Status: 12 characters, ASCII Hex (12 bytes)
Sample measurement time [sec]: 1... 9999 (ushort) (2 bytes)
Nuclide name / Measurement type name 1: Text [7] (8 bytes)
Measured value 1: float (4 bytes)
Unit 1: Text [6] (7 bytes)
Integral value 1: float (4 bytes)
Integral unit 1: Text [6] (7 bytes)
Accuracy 1: Float (4 bytes)
Nuclide name 2: Text [7] (8 bytes)
Measured value 2: float (4 bytes)
Unit 2: Text [6] (7 bytes)
Integral value 2: float (4 bytes)
Integral unit 2: Text [6] (7 bytes)
Accuracy 2: Float (4 bytes)
Integral time since last reset [h]: float (4 bytes)
Required disk space per record: (94 bytes)
The depth of the FIFO is 2400 measured data. With Read
FIFO, the above data are transferred in this order after sending
the header data (frame).
The current status of the measuring device is determined every
second. This includes the states of all errors and alarms. This
information is summarized in 12 Ascii hex characters, with each
bit corresponding to a particular state.
These 12 characters are transmitted with the Get status com-
mand and with Store automatically each current measurement
time is saved to memory.
On request, the communication protocol can be supplied with a
command list.
2.5 Scope of Delivery
Basic equipment Measurement and display unit LB 134 with integrated dose rate
probe LB 1346.
Standard accessories 1 set of Ni/MH rechargeable batteries (Mignon (AA) 1.2V/2.5Ah)
1 power supply unit (PSU)
1 bag with strap
User manual
Optional accessories wall bracket for LB 134 basic unit (incl. screws and dowels)
Wall bracket for LB 1343 contamination probe
Different probes for the desired measurement tasks