LB 134 7. Probes for the LB 134
LB 1342:118mm x 145mm, 170 cm²
LB 1343: 150 mm x 230 mm, 300 cm²
Weight per unit area: 0.4 mg/cm²
LB 1342: Alpha channel: approx. 0.1cps,
beta channel: approx. 10cps
LB 1343: Alpha channel: approx. 0.1cps,
beta channel: approx. 15cps
Sensitivity to 1
µSv/h external
gamma radiation
(Cs 137)
Alpha channel: not detectable
Alpha into beta channel (Po-210): < 20%
Beta into alpha channel (Sr-90): < 2*10
-5
Measuring range
(linearity devia-
tion error
< 10%)
Alpha channel: 0 – 5 000cps
Beta channel: 0 – 50 000cps (for alpha channel
< 750cps)
Temperature range operation: -20°C ... +40°C
(no condensation)
Stability of background over T-range: 10%
Temperature range storage: -40°C ... +60°C
β: >50000cps Overrange Indication
α: >5000cps Overrange Indication
Light-proof/
Entrance
window
Close to PC monitor no visible influences