Protocols
36
Figure 36 shows a melt curve step added after step 6:
Figure 36. Protocol with inserted melt curve step.
Step Options
To change a step option for the selected step:
1. Select a step by clicking on the step in the graphic or text view.
2. Click the Step Options button to open the Step Options window.
3. Add or remove options by entering a number, editing a number, or clicking a check box.
TIP: To hold a step forever (an infinite hold), enter zero (0.00) for the time.
Figure 37 shows the selected step with a gradient of 10
o
C. Notice that some options are not
available in a gradient step. A gradient step cannot include an increment or ramp rate change.
Figure 37. Step option for a gradient.
NOTE: A gradient runs with the lowest temperature in the front of the block (row H)
and the highest temperature in the back of the block (row A).
The Step Options window lists the following options you can add or remove from steps:
Plate Read. Check the box to include a plate read
Temperature. Enter a target temperature for the selected step
Gradient. Enter a gradient range for the step

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BIO RAD CFX96 and is the answer not in the manual?

BIO RAD CFX96 Specifications

General IconGeneral
Temperature Uniformity±0.4 °C
SoftwareCFX Maestro Software
Detection MethodReal-time PCR (qPCR)
Sample Capacity96-well plate
Excitation SourceLED
Ramp RateUp to 5°C/sec
Heating Rate5°C/sec
SensitivitySingle copy detection
Power Requirements100-240 V, 50/60 Hz
Temperature Range0–100 °C

Related product manuals