6.4.8 SpO
2
Signal Lost Alarm
Property Description
Alarm text
SPO2 Signal Lost
Alarm condition SpO
2
signal lost.
Priority High
Possible cause Signal lost reported by SpO
2
electronics (due to patient remov-
ing the probe from finger, or sensor detached from SpO
2
electronics.
Ventilator action The ventilator will continue treatment with the same settings.
Reset
User presses OK or electronics cable is disconnected by the
user, or the sensor is reconnected to the finger.
6.4.9 Poor SpO
2
Signal
Property Description
Alarm text
Poor SPO2 Signal
Alarm condition A Poor SpO
2
signal alarm will be given when the SpO
2
signal is
not correct. Check the SpO
2
sensor.
Priority High
Possible cause Artifact or low perfusion reported by SpO
2
electronics
Ventilator action The ventilator will continue treatment with the same settings.
Reset
OK message from SpO
2
electronics or SpO
2
electronics discon-
nected by user or SpO
2
Signal Lost alarm is triggered.
98
Alarms
User Manual
Doc. 006553 M-6