6.3.5 High EPAP Alarm
Property Description
Alarm text
High EPAP
Priority
Medium
Alarm condition A High EPAP alarm will be given when the measured EPAP is
30% above the set value for more than 15 seconds
Possible cause
Blocked leakage port.
Too short expiratory time.
Changes in airway resistance and or compliance.
Reset criteria
EPAP has gone below the alarm limit (lower than 30% above
the set value).
Ventilator action The Nippy 4 will continue treatment according to the current
settings.
Setting range
On
Off
6.3.6 Low EPAP Alarm
Property Description
Alarm text
Low EPAP
Priority
Medium
Alarm condition A Low EPAP alarm will be given when the measured EPAP is
30% below the set value for more than 60 seconds
Possible cause
Excessive leakage.
Reset criteria
EPAP has gone above the alarm limit (higher than 30% below
the set value).
Ventilator action The Nippy 4 will continue treatment according to the current
settings.
Setting range
On
Off
Alarms
85
Doc. 006553 M-6 User Manual