4 Preparing the Nippy 4 for Use
WARNING!
Read 2 Safety Information, page 14 before setting up the Nippy 4.
4.1 Checking the Nippy 4 before First Use
When using the Nippy 4 for the first time, follow the instructions below:
1 Check that all main components and
ordered accessories have been delivered
(refer to the packing note or the invoice, if
available).
2 Ensure that the equipment is in good condition.
3 If stored more than 1 month, connect the Nippy 4 to the power supply to recharge the
internal battery.
1
2
4 Check that the grey and white air filters
are installed.
4.2 Placing the Nippy 4
WARNING!
Read 2.3 Environmental Conditions, page 17 carefully to make sure all conditions are met
and considered.
1 Place the Nippy 4 on a solid, flat, and
clean surface.
The Nippy 4 should be placed lower than
the patient in order to prevent the device
from falling on the patient, as well as
preventing condensed water from reaching
the patient.
Overnight, the Nippy 4 should be placed
close enough to the patient’s bedside to
allow movements during the sleep without
pulling the Nippy 4 of its surface.
Preparing the Nippy 4 for Use
35
Doc. 006553 M-6 User Manual