6.4.17 Internal/Click-In Battery Hot Alarm
Property Description
Alarm text
Internal Battery — Internal Battery Hot
Click-In Battery — Click-In Battery Hot
Alarm condition
An alarm for Internal/Click-In Battery Overheat in Discharge
will be given when the internal or click-in battery reaches 55°C
(131°F).
Priority High
Ventilator action The ventilator will continue treatment with the same settings.
Battery discharging will be disabled (by the battery electronics)
once the temperature gets to 60°C (140°F). (If the battery is last
power source, the ventilator will stop running).
NOTE
The battery electronics by manufacture stops discharge at 60°C (140°F).
6.4.18 Heated Circuit Temp. Alarm
Property Description
Alarm text
Heated Circuit Temp.
Alarm condition A Heated Circuit temp alarm will be given when the measured
temperature of the heated wire is outside the tolerance.
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
Priority Medium
Reset
Heated wire measured temp tolerance is inside limits.